6-20
Pausing a Run
Once a run has started, you can manually pause a run, in order
to add a reagent to your samples, for example, using the PAUS E
key. Note that the instrument will automatically pause if you
select the Pause option in a CYCL or AUTO program.
! WARNING !
Be careful not to touch the sample block or the
bottom of the heated cover during a pause. These areas may be
hotter than 100 degrees C.
1
Press the
PAUS E
key.
The instrument beeps and a display similar to the one shown below
appears.
This display occurs in a CYCL program.
2
Press
RUN
to resume the run (from where you left off).
Stopping a Run
1
To abort a run at any time, press the
STOP
key.
The following appears on the second line of the display:
2
If you want to abort the run, press
ENTER
to accept the YES
selection.
OR
CYCL# 26 xx.xC
PAUSE 10:00
Abort program? YES