8-5
2
Using a cotton swab, coat wells D1 and E1 with mineral oil.
Figure 8-1 shows the location of these wells.
Figure 8-1
Location of sample wells E1 and D1.
3
Place the probe tray on the sample block so that the probe tray
notch faces the front of the instrument.
The RTD probe assembly consists of two cones, one of which
measures the temperature of the sample well. The wire is attached
to the cone that does not measure the temperature of the sample
well. This cone is a dummy probe.
4
Place the probe assembly into wells D1 and E1 so that the dummy
probe sits in D1. Thread the probe wire through the notch in the
probe tray (see Figure 8-2). Make sure the probe is connected to
the digital thermometer.