8-9
When the clock reaches three minutes, the following display
appears:
13
Measure the temperature of well E1 using the digital thermometer
and record this temperature as T(95).
Note
To exit the test at any time, press the
STOP
key. Press
5
and
ENTER
to return to the Verify Calibration Diagnostic test.
Figure 8-3 shows when you should take the two measurements.
Figure 8-3
Temperature measurements in the temperature calibration
verification test.
Record Temperature
Time=MM:SS Blk=95.0C
95
40
Tem p
°C
Time
0
3
0
3
Clock
starts
Measure
well E1 T(40)
Clock
starts
Measure
well E1
T(95)